Kryo 750-30 Features & Technical Specifications
The Kryo 750-30 is a flexible cryogenic system designed for
freezing of Bone Marrow, Stem Cells, Pharmaceutical Cell Lines, Skin, Cord Blood and other critical samples in high volumes. It is
designed for freezing of samples in bags, ampoules and straws.
It's 2 button operation is unique and it's standard PC software enables password protected multiple protocols. The Protocol stage trigger on sample, chamber temperature, or time. The unique forced laminar flow cooling system ensures most efficient, even cooling. The Top, or Front opening provides for easy loading, and the heated door seal prevents freezing shut at cryogenic temperatures. The Inner chamber removable for sterilization
Standard operating features include:
Start above ambient
Controlled heating
Comms port for PC connection
Fast cooling rates
TECHNICAL SPECIFICATION
Dimensions
Front Loading
External Internal
Height 55cm 26cm
Width 79cm 46.5cm
Depth 48cm 25cm
Top Loading
External Internal
Height 48cm 25cm
Width 79cm 46.5cm
Depth 55cm 26cm
Weight 45kg (shipping weight including packaging) approx.
Capacity
1452 x 2ml ampoules, or
1216 x 0.5ml straws (horizontal or vertical), or
1216 x 0.25ml straws (horizontal or vertical), or
40 x 50ml blood bags, or
20 x 250ml blood bags, or
20 x 500ml blood bags
Circulation Horizontal laminar flow
Temperature range +100.0șC to -160șC
Cooling medium Liquid nitrogen 22 ±2 psi.
Heater 1000W.
Sensors Control and sample: 4-wire Platinum resistance thermometer
Accuracy
±0.5șC at a hold at 0șC (dynamic accuracy depends on actual program, e.g. rate of change of temperature)
Heating rates 0.01șC/min to 10șC/min.
Cooling rates -0.01șC/min to -10șC/min.
Programmable Cooling Rate Range -0.01șC/min to -99.9șC/min.
Operating positions Vertical or horizontal
Thermal cutout 120șC cutout
Power 103-126VAC 50/60Hz 1200VA (max.)
Chart Sensitivity: 16.7mV/șC. Nominal impedance >10K
Recorder Scaling: OV = -200șC, +5V = +100șC
Standards
Designed to comply with BSEN 61010,
CSA22.2No.125-M1984,
CSA22.2No.151-M1986, EN50082-2,
EN50081-2.
Storage temperature -10șC to +70șC
Storage humidity Up to 95% noncondensing
Operating temperature 5șC to +40șC
Operating humidity Less than 90%